Pre-Set Signal Outputs - Weinmann WM 24037 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Signal
Pressure
Flow
Leakage
Pressure
Flow
rRMV
Leakage
Status

3.6 Pre-set signal outputs

The following signals are pre-set for display in WEINMANNsupport.
SOMNOsmart
SOMNOsmart 2
SOMNOset
SOMNOsoft +
SOMNObalance (e)/
SOMNOsoft 2 (e)
SOMNOvent CR
SOMNOvent auto-S,
SOMNOvent auto-ST
SOMNOvent CR
0 - 1 V DC ≡ 0 - 20 hPa
(up to and including firmware Version 4.0:
0 - 22 hPa)
0 - 1 V DC ≡ -250 - +250 l/min
0 - 1 V DC ≡ 0 - 100 l/min
SOMNOvent auto-S, SOMNOvent auto-ST
0 - 1 V DC ≡ 0 - 20 hPa
0 - 1 V DC ≡ -250 - +250 l/min
0 - 1 V DC ≡ 0 - 255 %
0 - 1 V DC ≡ 0 - 100 l/min
0 - 1 V DC ≡ 0 - 3
Analog channel 1
OPS
Leakage (up to
serial no. 2411:
pressure)
Leakage
Connecting lead PSG
Connect PSG connecting lead
Values
Analog channel 2
Pressure
Flow
EN
43

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Wm 23904Wm 23908Wm 23929Wm 23976Wm 23941

Table des Matières