14. Fault

Fault type
When pressing the On/Off
button, the button outline
flashes briefly in red and
the screen is not lit.
The battery level display
on screen indicates a low
level.
The screen is off but the
camera head is on.
Image in black and white
Bad visibility
The camera and/or screen
do not respond.
Blue screen indicating NO
SIGNAL on screen
Possible cause
Battery discharged.
Battery discharged.
The screen plug is not
inserted properly in
the outlet.
The colour intensity
setting is on
minimum.
The plastic dome is
scratched.
Battery discharged.
The device is faulty.
The flexible rod is
broken.
The screen's plug-in
connector is not
connected properly
The cables located in
the connector are
faulty.
Solution
Recharge the battery or connect the camera
to mains.
Recharge the battery or connect the camera
to mains.
Push the screen plug.
Move key 4 up to adjust colour intensity
▲.
Replace the plastic dome.
Recharge the battery or connect the camera
to mains.
Ship the device to the repair workshops.
Ship the device to the repair workshops.
Connect the plug-in connectors properly.
Ship the device to the repair workshops.