1. Insert the test strip into the test slot of the device. Wait for
the device to display the test strip "
" ".
2. The meter will detect the difference between control solution
and blood samples automatically. It will automatically mark
the result as a control solution test with "QC" display.
3. Apply the control solution. Shake the control solution vial
thoroughly before use. Squeeze out a drop and wipe it off,
then squeeze another drop and place it on the tip of the
vial cap. Hold the device to move the absorbent hole of the
test strip to touch the drop. Once the confirmation window
is filled completely, the device will begin counting down.
" and blood drop
EN-12
Note:
To avoid contaminating the control solution, do not directly
apply the control solution onto a strip.
4. Read and compare the result. After counting down to 0, the
test result of the control solution will appear on the display.
Compare this result with the range printed on the test strip
vial or individual foil pack and it should fall within this
range. If the test result is out of range, read the instructions
again and repeat the control solution test.
Note:
• Control solution test results are stored in the memory.
• The control solution range printed on the test strip vial or
individual foil pack is for control solution use only. It is not
the recommended range or reference values.
• Refer to the Maintenance section for important
information about your control solutions.
Out-of-range results:
If you continue to get results that fall outside the range
118~160