4
Temperature tables
Measuring systemPromag 53 W/P/H (compact version)
at T
= 40 °C
a
Promag H
DN 2...100
Promag P
DN 25...200
Promag P
DN 15...600
Promag W
DN 65...1200
at T
= 45 °C
a
Promag H
DN 2...100
Promag P
DN 25...200
Promag P
DN 15...600
Promag W
DN 65...1200
at T
= 50 °C
a
Promag H
DN 2...100
DN 25...200
Promag P
Promag P
DN 15...600
Promag W
DN 65...1200
Sensor Promag W/P/H (remote version)
at T
= 50 °C
a
Promag H
DN 2...100
Promag P
DN 25...200
Promag P
DN 15...600
Promag W
DN 65...1200
* The IP 68 version is limited to 130 °C.
at T
= 60 °C
a
Promag H
DN 2...25
Promag H
DN 40...100
Promag P
DN 25...200
Promag P
DN 15...600
Promag W
DN 65...1200
T6
85 °C
80
(PFA lining)
80
(PTFE lining)
80
(hard rubber lining)
80
T6
85 °C
80
(PFA lining)
80
(PTFE lining)
80
(hard rubber lining)
80
T6
85 °C
80
(PFA lining)
80
(PTFE lining)
80
(hard rubber lining)
80
T6
85 °C
80
(PFA lining)
80
(PTFE lining)
80
(hard rubber lining)
80
T6
85 °C
80
80
(PFA lining)
80
(PTFE lining)
80
(hard rubber lining)
80
Proline Promag 53
Max. medium temperature [°C] in
T5
T4
T3
100 °C
135 °C
200 °C
300 °C
95
130
150
150
95
130
150
150
95
130
130
130
80
80
80
Max. medium temperature [°C] in
T5
T4
T3
100 °C
135 °C
200 °C
300 °C
95
130
130
95
130
130
95
130
130
80
80
80
Max. medium temperature [°C] in
T5
T4
T3
100 °C
135 °C
200 °C
300 °C
95
95
95
95
95
95
95
95
95
80
80
80
Max. medium temperature [°C] in
T5
T4
T3
100 °C
135 °C
200 °C
300 °C
95
130
150
150
95
130
150 *
150 *
95
130
130
130
80
80
80
Max. medium temperature [°C] in
T5
T4
T3
100 °C
135 °C
200 °C
300 °C
95
130
130
130
95
130
150
150
95
130
130
130
95
130
130
130
80
80
80
Endress+Hauser
T2
T1
450 °C
150
150
130
80
80
T2
T1
450 °C
130
130
130
130
130
130
80
80
T2
T1
450 °C
95
95
95
95
95
95
80
80
T2
T1
450 °C
150
150 *
130
80
80
T2
T1
450 °C
130
150
130
130
80
80