KEY TO DISPLAY ICONS
Icon displayed
If any of the following marks is displayed, take the specified action, then
repeat the operation.
Icon displayed
See page 15.
See page 15.
Lit
Blinking
"Er" is displayed when the pulse value
is below 40 or above 181.
Condition/Cause
• Appears when the
blood-pressure could
not be measured
accurately.
The pressure is 281 mmHg
or higher.
Appears when the
batteries are running low.
Detected problem in the
device.
Eng 20
-
-
Condition/Cause
Measurement is under way.
Pulse detected.
Corrective action
Apply cuff properly and
measure again.
If the pressure does not
automatically decline
during measurement,
immediately press the
"START/STOP" switch
and turn the unit off.
Take the measurement
again.
Replace the batteries.
Contact the vendor.