3 Start of Application

Infrared data from the instrument will be received by Data Transfer at 9600
baud. The baud rate can be changed by the supplied software (see instruction
manual "Spectroquant
"ON" the last selected interface will be active.

3.1 PC interface mode

Press the "SELECT" key to scroll until the PC interface is set.
For data receiving on the PC a Data Transfer Software is required. For installation
and operation see "Spectroquant

3.1.1 Connection details

Connect the PC and the Data Transfer device port using a USB cable.

3.1.2 Data transmission

Keep the Data Transfer and the transmitting instrument (e. g. Spectroquant
Move 100 colorimeter) in line of sight (face to face, see figure 'Data Transfer –
Instrument connection' ). Start print command at the transmitting instrument.
The data will be sent to the Data Transfer and transmitted to the PC.

3.2 USB Printer interface mode

If necessary press the SELECT key to scroll until the USB interface mode is
selected. If the USB printer is not connected the interface status LED will flash.
If data arrives while the device is not ready, it will be lost.

3.2.1 Connection to USB printer

Connect the USB printer cable to USB host port on Data Transfer. If the printer
is successfully connected and ready to print the LED in front of the USB printer
port will illuminate.
Note:
The Data Transfer works in combination with printers with HPPCL up
to version 5, e. g. HP Deskjet 6940.
3.2.2 Data transmission
Keep the Data Transfer and the transmitting instrument (e. g. Spectroquant
Move 100 colorimeter) in line of sight (face to face, see figure 'Data Transfer –
Instrument connection' ). Start print command at the transmitting instrument.
The data will be sent to the Data Transfer and will be printed on the USB printer.
Note: The Data Transfer does not support and will not indicate any printer
related errors (Paper jam, paper tray empty, etc.).