To increase the user's safety, we recommend placing the bed in its lowest
position, especially when sleeping.
2.2.

Permitted loads

Maximum safe working load (including mattress, user, attachable
accessories)
Maximum occupant/patient weight
Maximum load on the backrest and foot section (each)
The maximum safe working load requires a weight distribution of approx. 45%
on the backrest, 25% on the middle section and 30% on the foot section.
2.3.

Ambient conditions

Equipment that generates strong magnetic fields which could interfere with the
bed's control unit should not be used in the bed's immediate vicinity. Con-
versely, the bed may generate electromagnetic fields during operation that
may affect nearby measuring equipment; nevertheless the bed complies with
the EN 60601-1-2 standard. Some residual risks remain.
The bed must not be used if there is a risk of explosion or in the presence of
flammable, volatile gases, e.g.anaesthetics.
The bed may only be used under the following ambient conditions:
- Temperature:
- Humidity:
- Air pressure:
In case of higher temperatures during delivery, let the bed rest for 24 hours
before using it or connecting it to the mains.
between 5°C and 40°C
20% to 90% at 30°C without condensation
700 to 1060 hPa
58
200 kg
150 kg
80 kg