Figure 7
3.3 Tungsten lamp on/off (W Lamp On/Off) (Fig. 8)
Note: For 4211/50, Zuzi recommends to have the tungsten lamp switched off when working out of the visible
range.
At the function selection display, move the cursor with
Confirm OK with F1 (2.5) to enter the tungsten lamp state selection display.
With
(2.7) (2.8) select On or Off and confirm OK with F1 (2.5). Beside the selected state the symbol
be displayed.
Confirm Cancel with F2 (2.6) to return to the function selection display.
Figure 8
3.4 Dark Current (0%T) (Fig. 9)
Zuzi recommends to carry out this function whenever changing the environmental conditions (changes in tem-
perature or equipment location, peaks in electrical current, etc.). For this purpose:
At the function selection display, move the cursor with
Confirm OK with F1 (2.5); Get dark current will be displayed.
After a few seconds, the equipment will automatically return to the function selection display and will be ready
to operate.
Figure 9
3.5 Wavelength calibration (Reset WL) (Fig. 10)
At the function selection display, move the cursor with
Confirm OK with F1 (2.5); Calibrating WL... will be displayed.
After a few seconds, the equipment will automatically return to the function selection display and will be ready
to operate.
Pág. 22
(2.7) (2.8) up to the W Lamp function.
(2.7) (2.8) up to the Dark current function.
(2.7) (2.8) up to the Reset WL function.
Instruction manual 542110x0
will
Version 2 Nov-10