Neither the manufacturer nor the distributor will assume any responsibility for the damages produced to the
equipment during its installation or damages to persons suffered by the improper use of the electric connection.
The tension should be 220 V, 50 Hz ± 10%.
Setting up
Press On/Off (1.4).
The equipment initializes and makes an autotest. This porcess can spend a few minutes; during this time Zuzi self
test... will be displayed. After this time, Warm up 20 minutes any key to skip will be displayed; wait for 20 minutes
before operating with the equipment.
NOTE: Do not open the sample compartment during autotest.
BASIC FUNCTIONS
1. Wavelength adjustment (Fig. 3)
λ
Press GOTO
(2.1) to enter the wavelength adjustment display. Select the desired value of wavelength by pres-
sing
(2.7) (2.8).
Select OK with F1 (2.5) to save the wavelength value.
The equipment automatically returns to the main menu.
If there is no need of changing the wavelength value, select Return with F2 (2.6) to go to the main menu.
Figure 3
2. Blank (Fig. 4)
Put the cell with the reference sample into the cell holder compartment.
Press ZERO (2.2); the equipment will make the blank automatically.
Figure 4
Pág. 20
100.0% T
100.0% T
Instruction manual 542110x0
100.0% T
100.0% T
Version 2 Nov-10