English
EBI16 checking the
production cycle control
Starting conditions
- EBI16 must be started at room
temperature
(+15°C
Place in the chamber
- If it is necessary, start before the
actual sterilization program a vacuum
test and/or a B&D test.
- Place the EBI16 in the Chamber. The
EBI16 must be free without any
obstruction. Do not place it inside a
container or similar.
- Place the EBI16 vertical in the
chamber.
- Make sure that the EBI16 is not in the
area of the steam entry.
- Start the sterilizer immediately after
placement of the EBI16.
Evaluation
- Interface and software are required
for the operation of the EBI16
- Automatic
evaluation
production cycle control can be done
only
with
Winlog.validation.
Archiving of the results
- The archiving of
ensured permanently by Winlog.med
or Winlog.validation.
ba80000def01_EBI16
...
+35°C).
of
the
Winlog.med
the
results
Français
Effectuer un contrôle de cycle
de routine avec l'EBI 16
Conditions de depart
- L'EBI
16
température
(+15°C...+35°C)
Positionnement dans la chambre
- Si nécessaire, commencer avant le
programme de stérilisation actuel un
test du vide ou un test de B&D
- Positionner l'EBI 16 dans la chambre.
Aucun obstacle ne doit se trouver
devant l'EBI 16. Ne pas le positionner
à l'intérieur d'un récipient ou autre.
- Positionner l'EBI 16 verticalement
dans la chambre.
- Veuillez vous assurer que l'EBI 16 ne
se trouve pas à l'endroit où pénètre la
vapeur.
- Démarrer
immédiatement
positionnement de l'EBI 16.
Evaluation
- Une interface et un logiciel sont
nécessaires pour utiliser les EBI 16.
- L'évaluation automatique du test de
B&D peut être effectué seulement
or
avec les logiciels Winlog.med ou
Winlog.validation.
Archivage des résultats
- L'archivage des résultats est assure
is
en
permanence
Winlog.med ou Winlog.validation
doit
être
démarré
ambiante
le
stérilisateur
après
par
le
à
le
logiciel
15