EBI16 checking the Vacuum
test

Starting conditions

- The start temperature is not important
for vacuum test.

Place in the chamber

- Place the EBI16 in the center of the
Chamber.
- Place the EBI16 vertical in the
chamber.
- Make sure that the EBI16 is not in the
area of the steam entry.
- Start the sterilizer immediately after
placement of the EBI16
- During the vacuum test, no other load
is allowed in the chamber.

Evaluation

- Interface and software are required
for the operation of the EBI16
- Automatic evaluation of the vacuum
test can only be carried out with
Winlog.med
or

Archiving of the results

- The archiving of
ensured permanently by Winlog.med
or Winlog.validation.
ba80000def01_EBI16
Winlog.validation.
the
results
is
Français
Effectuer le test du vide avec
l'EBI 16

Conditions de depart

- La temperature de départ n'est pas
importante pour le test du vide.

Positionnement dans la chamber

- Positionner l'EBI 16 au centre de la
chambre
- Positionner l'EBI 16 verticalement
dans la chambre
- Veuillez vous assurer que l'EBI 16 ne
se trouve pas à l'endroit où pénètre la
vapeur
- Démarrer
immédiatement
positionné l'EBI 16
- Pendant le test du vide, aucune autre
charge
n'est
chambre
Evaluation
- Une interface et un logiciel sont
nécessaires pour utiliser l'EBI 16
- L'évaluation automatique du test du
vide peut être seulement effectuée
avec
le
logiciel
Winlog.validation
Archivage des résultats
- L'archivage des résultats est assuré
en permanence par les logiciels
Winlog.med ou Winlog.validation
le
stérilisateur
après
avoir
autorisée
dans
Winlog.med
la
ou
11

Publicité

Table des Matières
loading

Table des Matières