EBI16 checking the Vacuum
test
Starting conditions
- The start temperature is not important
for vacuum test.
Place in the chamber
- Place the EBI16 in the center of the
Chamber.
- Place the EBI16 vertical in the
chamber.
- Make sure that the EBI16 is not in the
area of the steam entry.
- Start the sterilizer immediately after
placement of the EBI16
- During the vacuum test, no other load
is allowed in the chamber.
Evaluation
- Interface and software are required
for the operation of the EBI16
- Automatic evaluation of the vacuum
test can only be carried out with
Winlog.med
or
Archiving of the results
- The archiving of
ensured permanently by Winlog.med
or Winlog.validation.
ba80000def01_EBI16
Winlog.validation.
the
results
is
Français
Effectuer le test du vide avec
l'EBI 16
Conditions de depart
- La temperature de départ n'est pas
importante pour le test du vide.
Positionnement dans la chamber
- Positionner l'EBI 16 au centre de la
chambre
- Positionner l'EBI 16 verticalement
dans la chambre
- Veuillez vous assurer que l'EBI 16 ne
se trouve pas à l'endroit où pénètre la
vapeur
- Démarrer
immédiatement
positionné l'EBI 16
- Pendant le test du vide, aucune autre
charge
n'est
chambre
Evaluation
- Une interface et un logiciel sont
nécessaires pour utiliser l'EBI 16
- L'évaluation automatique du test du
vide peut être seulement effectuée
avec
le
logiciel
Winlog.validation
Archivage des résultats
- L'archivage des résultats est assuré
en permanence par les logiciels
Winlog.med ou Winlog.validation
le
stérilisateur
après
avoir
autorisée
dans
Winlog.med
la
ou
11