Technical alarm – Description
If the (systolic or diastolic) blood pressure taken is
outside the range indicated in the "Technical Fea-
tures" paragraph, the technical alarm displays the
message "Hi" or "Lo" on the screen. In this case,
consult a doctor or check that you are using the device
properly. The technical alarm's limit values are values
set by the factory and can neither be changed nor
deactivated. Under the IEC 60601-1-8, these limit
values of the alarm have a lower priority. The tech-
nical alarm is not a locked alarm and does not need to
be reset. The signal displayed on the screen switches
off automatically after approximately 8 seconds.
MAINTENANCE AND STORAGE
•
Clean the device and the cuff carefully, using only
a slightly damp cloth.
•
Do not use detergent or solvent.
•
Never put either the device or the cuff in water
as this might seep into the device or cuff and
damage them.
•
Do not put heavy objects on the device or on
the cuff when they are being stored. Remove
the batteries.
•
No component can be maintained by user in the
monitor. The circuit diagrams, component part
lists, descriptions, calibration instructions, or
other information which will assist the user's
appropriately qualified technical personnel to
repair those parts of equipment which are des-
ignated repairably can be supplied.
•
It is recommended the performance should be
checked every 2 years or after repair. Please
contact the service center.
•
The monitor can maintain the safety and perfor-
mance characteristics for a minimum of 10,000
measurements or three years.
•
The cuff integrity is maintained after 1,000
open–close cycles of the closure.
•
It is recommended the cuff should be disinfect-
ed 2 times every week if needed (For example,
in hospital or in clinique). Wipe the inner side
(the side contacts skin) of the cuff by a soft cloth
squeezed after moistened with Ethyl alcohol
(75-90%), then dry the cuff by airing.
•
Clean the cuff after the usage of every 200 times
is recommended.
TECHNICAL SPECIFICATIONS
OF THE PRODUCT
Model number LA090206
Measuring
Measuring blood pressure on the arm,
method
oscillometric and non-invasive
Measuring
Pressure in the cuff 0 – 300 mmHg
range
Systolic pressure 60 – 260 mmHg
Diastolic pressure 40 – 199 mmHg
Pulse 40 – 180 beats/minute
Accuracy of
systolic ±3 mmHg
the indicator
diastolic ±3 mmHg
pulse ±5% of the value displayed
Memory
4 x 30 memory locations
Dimensions
+/- 107 x 80 x 52mm
Weight
+/- 287g (without any batteries)
Cuff size
22 - 30cm
Admin. operat-
+ 10°C a + 40°C, ≤85% relative air
ing conditions
humidity (non-condensing)
Permissible
-20°C a +50°C, ≤85% relative air humidity,
storage
800 – 1 050 hPa of ambient pressure
conditions
Power supply
4 x 1.5 V AAA batteries
Battery lifespan Approximately 200 measurements,
depending on the increased blood
pressure and inflation pressure
Accessories
Instructions for use, storage box
Ranking
Internal power supply, IP20, no AP
or APG, continuous use, BF device
Pressure meas-
Pump,Valve, LCD, Cuff, Sensor
uring system
components
Changes could be made to the technical features
without any prior warning for updating purposes.
•
This device complies with European standard
EN60601-1-2 and meets the special safety re-
quirements concerning electromagnetic compat-
ibility. Please note that HF portable and mobile
communication devices are likely to affect this
device. For further details, please contact the
after-sales service at the address stated or consult
the end of the instructions for use.