8. DISINFECTION AND STERILIZATION
8.1 DISINFECTING THE UNIT
NOTE
Always disconnect the power cord before cleaning the system
Use any disinfectant agents which are commonly applied while disinfecting surfaces of
electric medical equipment. Such disinfectant agents usually arrive in the form of
sprays or damp cloths.
Follow the instructions given by the manufacturer of the disinfectant solution.
8.2 STERILIZING THE LIGHT CABLE
The light cable should be sterilized according to the sterilizer manufacturer's
specification.
9.
MAINTENANCE, SERVICING AND REPAIR
Performance of preventive maintenance is not essential. Regular maintenance can,
however, contribute to identifying potential problems before they become serious; thus,
enhancing the instrument's reliability and extending its useful operating life.
Maintenance services can be obtained from your local representative or from the
manufacturer.
Defective items of equipment are to be serviced and repaired exclusively by persons
authorized by the manufacturer. All repair work shall employ original manufacturer's
parts only.
9.1 LAMP REPLACEMENT
CAUTION Always disconnect power cord and turn main switch off before lamp
replacement.
Use lamp replacement P/N SYC0029
Turn light source off.
Allow light source to cool completely.
Disconnect power cord from outlet.
Turn locking screws counter clockwise and slide out lamp drawer as far as the stop.
To remove the lamp, push the lever toward the back of the unit, this will raise the lamp.
Remove the connector at the back of the lamp.
Replace the lamp with the new unit.
Slide the lamp drawer in and turn the locking screws clockwise until secured.
Insert power cord and turn power switch on.
®
LIT091 CUDA
SURGICAL
(English)
Rev. G
Page 10 of 60