EMC TABLES
Emissions test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
CEI 61000-3-2
Voltage variations /
Flicker emissions
CEI 61000-3-3
EMC TABLES
Immunity Test
Electrostatic Discharge
(DES)
CEI 61000-4-2
Power Supply
Frequency (50/60 Hz)
Magnetic Field
CEI 61000-4-8
Notice-Medicaid-tensio-858754-v1-042017-r1.indd 37
Manufacturer's declaration and directive – electromagnetic emissions
The 0081N wrist blood pressure monitor is intended for use in the electromagnetic environment
speci ed below. The client or user of the 0081N wrist blood pressure monitor must be sure to use it
in such an environment.
Compliance
The 0081N wrist blood pressure monitor uses RF energy only for its internal operation.
Group 1
Therefore, its RF emissions are very low and are not likely to cause interference in electronic
equipment nearby.
Class B
Not
The 0081N wrist blood pressure monitor may be used in all establishments, including
applicable
dwellings and establishments directly connected to the public low-voltage power supply
system that powers buildings intended for homes.
Not
applicable
Manufacturer's declaration and directive – electromagnetic immunity
The 0081N wrist blood pressure monitor is intended for use in the electromagnetic environment
speci ed below. The client or user of the 0081N wrist blood pressure monitor must be sure to use it
in such an environment.
Compliance
CEI 60601
Test level
level
± 6 kV
± 6 kV
The oors must be made of wood, cement or ceramic tiles. If the oors are
contact
contact
covered in synthetic material, relative humidity must be below 30%.
± 8 kV air
± 8 kV air
The magnetic eld of the supply frequency must be at a level appropriate
3 A/m
3 A/m
to the place of use.
Electromagnetic environment guidelines
Guidelines on the electromagnetic environment
37