4.3
Ratio between distance from target / diameter
of measurement field
This ratio, also called the field of vision, indicates the diameter S
of the measurement field at a given distance D from the target
(see figure below).
Diameter of measurement field : D/S = 8/1
Distance from target
In the case of small measurement points; it is important to
bring the probe sufficiently close to the target to avoid including
other points in the measurement field.

5. SPECIFICATIONS

Measurement range: -30 to +550°C
I I I I I
Resolution: depending on the multimeter
I I I I I
Accuracy: ±2% ±2°C of the reading
I I I I I
Spectral response: 6 to 14µm
I I I I I
Emissivity: preset at 0.95
I I I I I
Output voltage: 1mV/°C
I I I I I
Response time: 0,5sec
I I I I I
Utilisation environment
I I I I I
0 to +50°C < 80% RH
Storage environment
I I I I I
-20°C to +60°C , 0 to 80% RF, without battery
Power supply:9V battery
I I I I I
Dimensions / Weight
I I I I I
164 x 50 x 40mm / 182g with batteries
Electromagnetic compatibility
I I I I I
NF EN 61326-1 (Ed97) + A1