3-6 Rubber Pad
To prevent the control unit from slipping apply
rubber pads.
1 ) Clean bottom of the Control Unit.
2 ) Fit the Rubber Pad appropriate place as shown
on the Fig. 6.
Control Unit Holder is not necessary when rubber pads are applied.
•
CAUTION
4. Sterilization before first time use
NSK Varios components are not sold sterile and should be sterilized prior to first time you use.
Autoclave sterilization is recommended.
Please note the following when sterilize components.
Follow your local directives, standards, and guidelines.
Sterilize components in a U.S. FDA-cleared sterilization pouch, complaint to ISO 11607-1.
Ensure that you only remove dry sterile goods from the sterilizer.
Store sterile goods dust-free and dry.
• NSK recommends sterilization according to ANSI/AAMI ST79 and EN 13060 (class B).
NOTE
• Follow the manufacturer's instructions for use that came with your sterilizer.
• For more information about sterilization standards, refer the medical device regulations
specific to your location.
Autoclave Procedure
1) Pick a tip up from the tip case. Take handpiece out of the packing bag before sterilization.
2) Insert sterilizable components into a U.S. FDA-cleared sterilization pouch, complaint to ISO 11607-1. Seal the pouch.
3) Autoclave conditions are below:
< When sterilizing a Tip>
Type
Minimum Temperature
Full Cycle Time
Minimum Drying Time
* Not for use in the U.S.
<When sterilizing components except for a Tip>
Type
Minimum Temperature
Full Cycle Time
Minimum Dry Time
Steam Sterilization Cycle
Gravity Displacement
132°C
135°C
15 minutes
10 minutes
15 minutes
30 minutes
Steam Sterilization Cycle
Gravity Displacement
121˚C
132˚C
20 minutes
15 minutes
10 minutes
10 minutes
Rubber
Pad
Pre-Vacuum (Dynamic Air Removal)
132°C
134°C*
4 minutes
3 minutes
20 minutes
30 minutes
Pre-vacuum
135˚C
135˚C
10 minutes
10 minutes
10 minutes
15 minutes
8
135°C
3 minutes
16 minutes
Fig. 6