Proceed as follows to adjust the brightness of the ABU62 display
from the Titration Manager:
-
Display the Cell window.
-
At the line Instrument, select the ABU62 (ABU1 or ABU2) you want
to adjust the display brightness. ABU1 means an ABU62 connected
to the Local socket of the Titration Manager. ABU2 means an ABU62
connected to the Slave socket of a second ABU62 of the ABU1 type
(see figure on the previous page).
-
Press 0 to increase the brightness or 7 to decrease the brightness.
The ABU62 can only be operated from the Titration Manager ("Slave"
mode). It means that:
.
The Configuration parameters of the Titration Manager are imposed to
the ABU62 (language, date and time in particular),
.
The electrode and reagents used on the ABU62 are created and
edited from a Titration Manager screen,
.
The burettes, electrodes and reagents used on the ABU62 are
installed from a Titration Manager screen,
.
Direct measurements using electrodes connected to the ABU62 are
initiated on the Titration Manager,
.
Manual dosing using ABU62 burettes is initiated on the Titration
Manager.
All these operations are described in the Reference Manual of the Titration
Manager: consult the "Glossary" chapter at the appropriate keyword index
(examples: Create electrode, Electrode calibration, Manual dosing).