Sample Measurement; Conductivity Measurement; Tds, Salinity Or Resistivity Measurement - OHAUS STARTER 3100M Manuel D'instruction

Analyseur de table multi paramètre
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN-16
4.3

Sample measurement

Conductivity measurement

4.3.1
Place the conductivity electrode in the sample, wait 10-15s, then press
measurement.
The measurement icon
measurement and the display shows the conductivity of the sample.
When it reaches endpoint (auto endpoint or manual endpoint), you can record the result.
Stability criterion for conductivity measurement: The sensor input signal of the meter may not
change by more than 0.4% from the measured average conductivity of the electrode in 6 seconds.

TDS, Salinity or Resistivity measurement

4.3.2
Press
to switch between conductivity, TDS, Salinity and Resistivity measurement
mode.
Follow the same procedure as for conductivity measurement.
Ohaus conductivity electrodes STCON3 and STCON7 already have a built-in temperature sensor
to measure the temperature automatically.
appears on the display. The measurement icon is blinking during
STARTER 3100M Bench Meter
to start the

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

St3100m-nSt3100m-b

Table des Matières