Set Calibration Standard Point; Set Temperature Correction Coefficient; Set Tds Factor; Calibration - OHAUS STARTER 3100M Manuel D'instruction

Analyseur de table multi paramètre
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN-14
Press
to confirm your setting. Continue with calibration standard setting by following or
press
to go back to the measurement screen.

Set calibration standard point

4.1.3
After the MTC temperature value setting, the meter goes to setup calibration standard.
The 5 predefined standard points are:
10 µS/cm
84 µS/cm
Use the
or
Press
to confirm your setting. Continue with temperature correction coefficient setting
by following the steps in 4.1.4 or press

Set temperature correction coefficient

4.1.4
After the calibration standard setting, the meter goes to the temperature correction coefficient
setting.
Use the
or
Press
to confirm your setting. Continue with reference temperature setting by following
the steps or press
Note:
If you set the temperature correction coefficient value as 0 %/° C , that means no Temperature
Compensation for the conductivity measurement. The meter displays the real conductivity value
at current temperature without compensation.

Set TDS factor

4.1.5
After selecting the reference temperature, the current TDS factor blinks.
Use
or
setting or press
4.2

Calibration

The first time you use the conductivity electrode, it should be calibrated before taking any
measurement. Please also read the conductivity electrode instruction manual for reference.
500 µS/cm
buttons to select calibration standard.
buttons to adjust the temperature correction value. (0.00 to 10.00 %/° C).
to go back to the measurement screen.
buttons to adjust the TDS factor value. Press
to go back to the measurement screen.
STARTER 3100M Bench Meter
1413 µS/cm
12.88 mS/cm
to go back to the measurement screen.
to confirm your