Cleaning And Maintenance; Technical Specification; Troubleshooting; Metrological Control - Reer SoftTemp Instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22

Cleaning and maintenance

The sensor is the most sensitive part of the infra-red
thermometer. Please make sure you don't touch or scratch
the surface of the measuring sensor. If the sensor is dirty,
carefully clean it with a cotton swab soaked with alcohol or a
mild detergent (e.g. a highly diluted dish washing detergent).
Do not use any aggressive cleaning agents. Clean the case
and display with a soft dry cloth or an alcohol pad.
The device is not water-proof and thus must not be immer-
sed in water or exposed to direct humidity.
Remove the batteries from the device if it is not to be used
for lengthy periods of time.
This product does not require any special maintenance.
Repairs must be carried out by skilled personnel only.

Technical specification

Contactless infra-red technology
Measuring
technology
Measuring range
Body temperature (temple):
32,0°C – 42,9°C (89,6°F – 109,2°F)
Object/ambient temperature:
0°C – 100°C (32,0°F – 212,0°F)
Measuring accuracy
Body temperature: ± 0,2°C (0,4°F)
Object temperature: ± 0,3°C (0,5°F)
Ambient temperature: ± 1°C (1,8°F)
Celsius (°C) and Fahrenheit (°F)
Measuring units
Measuring distance
1 - 6 cm distance to measuring point
Measuring duration
1 second
19 measurements
Measured
value memory
Measured
Liquid crystal display (LCD)
value display
with backlighting
Green LED for ≤ 37,3°C (99,1°F)
LED display
Orange LED for 37,4°C – 37,9°C (99,2°F – 100,2°F)
Red LED for ≥ 38,0°C (100,3°F)
Acoustic signal
Can be switched on or off
30 seconds after last measurement
Automatic
switch-off
Size
162,5 x 46,5 x 42,5 mm
109,5 g (including batteries)
Weight
Batteries
2 x 1,5V AAA batteries (included in scope of delivery)
Protection degree
IP22
10°C – 40°C (60,8°F - 104°F) ambient temperature,
Operating
20% - 95% relative humidity,
conditions
70-106 KPa atmospheric pressure
Storing
-20°C – 50°C (-4°F - 122°F) ambient temperature,
conditions
≤ 95% rel. relative humidity,
70-106 Kpa atmospheric pressure
20

Troubleshooting

Problem/symbol
Possible cause
Device cannot be
Batteries are empty.
switched on
Batteries have been
inserted incorrectly.
Temperature
Sensor is dirty/defect
values wrong or
fluctuating
Low battery level,
measuring is still possible.
Batteries are empty, it is no
longer possible to measure
using this device.
The operating temperature
is outside the permitted
range of
10°C – 40°C (50°F – 104°F).
Measured body
temperature is over
42,9°C (109,2°F).
Measured object/
ambient temperature is
over 100°C (212°F).
Measured body
temperature is lower than
32°C (89,6°F).
Measured object/
ambient temperature is
lower than 0°C (32°F).

Metrological control

As a rule, an inspection of the thermometer is not required.
For devices used in medical practice, the German Medical
Products Operator Ordinance of 29 June 1998 requires a
metrological control after 2 years.
Measures
Insert new batteries.
Check position and polarity
of the batteries.
Check the measuring sen-
sor and clean it if necessary
as described on page 20.
Replace the
batteries soon.
Replace the batteries
immediately.
Take the thermometer
to a place where the
temperature is inside the
permissible range.
Acclimatise the thermome-
ter for at least one hour.
Check if you are in the
correct measuring mode
(body temperature/object
temperature).
Repeat the measurement.
Ensure the measuring sen-
sor is correctly positioned.
Check if you are in the
correct measuring mode
(body temperature/object
temperature).
Repeat the measurement.
Ensure the measuring sen-
sor is correctly positioned.
21