Technical Data - OHAUS Scout SJX Serie Guide De Démarrage Rapide

Masquer les pouces Voir aussi pour Scout SJX Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Quick Start Guide
LEGAL FOR TRADE (SJX...N/E models only)
When the balance is used in trade or a legally controlled application it must be set
up, verified and sealed in accordance with local weights and measures regulations.
It is the responsibility of the purchaser to ensure that all pertinent legal requirements
are met. As the requirements vary by jurisdiction, the purchaser is advised to
contact their local weights and measures office for instructions about putting the
balance into service.
Settings
Before verification and sealing, perform the following steps:
1.
Verify that the menu settings meet the local weights and measures
regulations.
2.
Units menu should be reviewed. Verify the units turned On are
permitted by the local weights and measures regulations.
3.
Perform a calibration.
4.
Set the position of the Security Switch to the locked position.
Verification
A weights and measures official must perform the verification procedure. Contact
the local weights and measures office for more information.
Sealing
After the Balance has been verified, it must be sealed by the weights and measures
official to prevent undetected access to the legally controlled settings. Before
sealing the device, ensure that the security switch is in the Locked position.
If using a wire seal, pass the sealing wire through the holes in the security switch
and Bottom Housing as shown.
If using a paper seal, place the seal over the security switch and Bottom Housing as
shown.
Unlocked

TECHNICAL DATA

Equipment Ratings:
Altitude: Up to 2000 m
Indoor use only
Operating temperature range: +5 ° C to +40 ° C
Specified temperature range: Refer to tables below
Relative humidity: 10% to 80% at 31° C, decreasing linearly to 50% at 40° C, non-
condensing
Power: AC power adaptor input 100-240V 50/60 Hz and output 5 V DC 1 A (For
use with certified or approved power supply, which must have a SELV and limited
energy circuit output.), or 4 AA batteries
Main supply voltage fluctuations: up to ± 10% of the nominal voltage
Pollution degree: 2
Installation category: II
Ohaus Corporation www.ohaus.com
Locked with Paper Seal
ix
TM
Scout
SJX Balances
Locked with Wire Seal
TM
Scout
SJX Balances

Publicité

Table des Matières
loading

Table des Matières