Calibration
The PosiTest AT is shipped with a Certificate of Calibration
showing traceability to a national standard. For organizations with
re-certification requirements, the PosiTest AT may be returned at
regular intervals for calibration. DeFelsko recommends that our
customers establish their instrument calibration intervals based
upon their own experience and work environment. Based on our
product knowledge, data and customer feedback, a one year
calibration interval from either the date of calibration, date of
purchase, or date of receipt is a typical starting point.
Verification
The PosiTest AT Verifier is available for verifying the accuracy and
operation of PosiTest Adhesion Testers and is an important
component in fulfilling both ISO and in-house quality control
requirements. Fully portable with hardshell carry case for use in
the field or laboratory. Learn more at: www.defelsko.com/at/verifier
Conforms to: ASTM D4541, ASTM D7234, ISO 4624 and others.
Specifications:
Resolution: 1 psi (0.01 MPa)
Dolly Size (mm)
10 mm
14 mm
20 mm
50 mm*
*requires the use of a 50 mm stand off
The instrument also measures in N/mm
force). See Units (pg. 12).
Calibration and Verification
Technical Data
Max Pull-Off
Pressure (AT-M)
10,000 psi (70 MPa) 14,000 psi (96 MPa)
6,000 psi (40 MPa)
3,000 psi (20 MPa)
500 psi (3.5 MPa)
16
Accuracy: ±1% Full Scale
Max Pull-Off
Pressure (AT-A)
7,000 psi (50 MPa)
3,500 psi (24 MPa)
560 psi (3.8 MPa)
and Newtons (unit of
2