PosiTest AT-A
Hold Time:
(up to 60 seconds)
pressure is
maintained (held)
at defined Limit.
Default: 0 seconds.
Orientation
Rotates display between landscape (default) or portrait views.
Ideal for use in horizontal or vertical positions.
Units
Converts the display from psi, MPa, N/mm
force). Existing measurements in memory are not converted.
Switching units closes Memory.
Gage Info
Displays the model number & serial number, PosiSoft.net
registration key, the amount of remaining memory for storage of
readings, date and time, and software packages.
For security purposes, the registration key is required to add the
Gage to your PosiSoft.net account (pg. 14).
Reset
Restores factory settings and returns the instrument to a known
condition. The following occurs:
- All batches, stored readings, notes, batch names and screen
captures are erased.
- Menu settings are returned to the following:
Memory = OFF
Pull Parameters = default
Orientation = Landscape
NOTE:
Date, Time and WiFi are not affected by a Reset.
In the unlikely event that the instrument becomes unresponsive
or will not power down, press and hold
screen fades to white, then release. The gage will restart.
Bluetooth = ON
WiFi = OFF
USB Drive = ON
12
Hold Time is
displayed on
the pull chart
(shaded area).
Defined
Hold Time
or Newtons (unit of
2
Units = psi
Backlight = Sun
Language = English
and
until the