Operating Procedure; In General - Chauvin Arnoux CA 10 Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9.1

IN GENERAL

Check that the resistance to be tested isn't live,
Switch ON the instrument by pressing the ON/OFF button (see 3 p.25),
Connect the resistance to be measured according to the diagram shown next to the
measurement terminals (see 1 p.25). With the instrument operating according to the
4 wire method, check the correct position of the voltage measurement points. To do
this, apply the voltage measurement instruments (the 10 A clamps supplied with the
instrument can be used both to input the measurement current and to measure the
voltage) at the precise points defining the resistance terminals.

9. OPERATING PROCEDURE

I
The value displayed by
the CA 10 MICROHMMETER
corresponds to the resistance
of the dotted part.
U
28
C 1
P 1
P 2
C 2