Operating Procedure; Switching On; Dc Zero Adjustment; Measurement - Chauvin Arnoux PAC 10 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

3/ OPERATING PROCEDURE

3.1/ SWITCHING ON

Set the sliding switch  to the position 400 A range (sensitivity 1 mV/A).
Correct operation is indicated by a green light  indicating that the battery
is in good condition.
If this green indicator does not come on when the clamp is switched on,
or goes out during the operation of the clamp, it is necessary to replace the
battery (see MAINTENANCE chapter).
3.2/ ZERO ADJUSTMENT
Ensure that the jaws  of the clamp are correctly closed and that they do
not enclose any conductor. Connect the clamp to your measurement
instrument set up according to the appropriate voltage range then turn the
rotative button “ until zero DC is displayed on the measurement instru-
ment

3.3/ MEASUREMENT

After having switched on the clamp, connected it to the measurement
instrument on the appropriate range, and set the zero DC procedure (see
the two paragraphs above), clamp the conductor to be measured Œ
between the jaws of the clamp.
The value is displayed on the measurement instrument.
Depending on the sensitivity selected on the clamp and the range of your
measurement instrument, apply the conversion ratio to get the value of
the current.
On DC current measurement, ensure that the arrow located on
the
external edge of the jaws  corresponds to the direction of the
current flowing in the conductor. (source ⇒ receiver)

4/ SPECIFICATIONS

Ranges
input/output
ratio
400 A
1 mV/A
4.1/ RÉFERENCE CONDITIONS
- Temperature: 18...28°C
- Humidity rate: 20...75% RH
- Battery voltage: 9 V ± 0.1 V
- Position of conductor: centred on the markings of the clamp
- Magnetic field: Earth's DC field
- Absence of external AC magnetic field
- Absence of electric field
- Measurement for a DC current or an AC sinusoidal current ≤ 65 Hz
- Impedance of the measurement instrument : ≥ 1 MΩ and ≤ 100 pF

4.2/ OPERATING CONDITIONS

The instrument must be used in the following conditions to satisfy the
safety of the user and the metrological performance:
- Use indoors
- Working altitude:≤ 2000 m
- Transportation altitude:≤ 12 000 m
- Environmental conditions: see graph below
measurement extent
AC rms
peak max
0.5...400 A
0.5...600 A
10
DC
0.5...600 A