Apex XT Fit Instructions D'utilisation page 22

Masquer les pouces Voir aussi pour XT Fit:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
XT Fit /English
a)Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the device is used exceeds the
applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and this device:
This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of this device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and this device as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
=
0.1
0.31
1
3.1
10
80 MHz to 800 MHz
=0.6
0.06
0.19
0.6
1.9
6
800 MHz to 2.7 GHz
=1.2
0.12
0.38
1.2
3.8
12
d
19
in

Publicité

Table des Matières
loading

Produits Connexes pour Apex XT Fit

Table des Matières