Apex XT Fit Instructions D'utilisation page 20

Masquer les pouces Voir aussi pour XT Fit:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
XT Fit /English
Appendix A: EMC Information
Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Warning:
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device
should be observed to verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the CPAP, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Immunity Test Levels
Professional
Basic EMC standard
healthcare
facility environment
Electrostatic
±8kV contact
Discharge (ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to earth
Compliance
Electromagnetic Environment-Guidance
Group1
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment
Class B
The device is suitable for use in all establishments, including domestic
Class A
establishments and those directly connected to the public low-voltage
power supply network
Complies
HOME HEALTHCARE
ENVIRONMENT
± 1 kV line(s) to
line(s)
Compliance
Electromagnetic
Levels
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
±8kV contact
with synthetic material, the
±15kV air
relative humidity should be at
least 30 %.
±2kV for power
Mains power quality should be
supply line
that of atypical commercial or
±1kV for
hospital environment
input/output line
± 1 kV line(s) to
Mains power quality should be
line(s)
that of atypical commercial or
hospital environment.
17

Publicité

Table des Matières
loading

Produits Connexes pour Apex XT Fit

Table des Matières