General Specifications - Klein Tools ET110 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ENGLISH
Klein Tools ET110 is an easy-to-use meter that detects and measures
concentration levels of carbon monoxide (CO), a colorless, odorless,
flavorless gas that is life-threatening even at relatively low concentrations.
The meter also measures temperature.
• CO Measurement/Detection Range: 0 to 1000 ppm
• Resolution: 1 ppm
• Repeatability: +/- 3%
• Accuracy:
(Test conditions: 40 ±10% relative humidity, 1 atm)
At 32° to 59°F (0 to 15°C): +/- 10% 10 ppm (whichever greater)
At 59° to 77°F (15 to 25°C): +/- 5% 5 ppm (whichever greater)
At 77° to 122°F (25 to 50°C): +/- 10% 10 ppm (whichever greater)
• Responding Time: T90 < 60 seconds
• Sensor: Electrochemical, replaceable
• Sensor Life Expectancy: 5 years
• Alarms: Low-Level: 35 ppm
High Level: 200 ppm
Short-Term Exposure Limit (STEL): 200 ppm for 15 minutes
• Temperature Range: 32° to 122°F (0° to 50°C)
• Temperature Resolution: 1°F/C
• Temperature Accuracy: +/- 5°F/C
• Batteries: 4x AAA
• Operating Altitude: 6562 ft. (2000 m)
• Relative Humidity: <85% non-condensing
• Operating Temp: 32° to 122°F (0° to 50°C)
• Storage Temp: 23° to 131°F (-5° to 55°C)
• Dimensions: 7.75" x 2.16" x 1.45" (197 x 55 x 37 mm)
• Weight: 6.87 oz. (195 g) including batteries
• Drop Protection: 3.3 ft. (1m)
Specifications subject to change.
To ensure safe operation and service of the meter, follow these
instructions. CARBON MONOXIDE IS LIFE-THREATENING EVEN AT
RELATIVELY LOW CONCENTRATIONS. Failure to observe these
warnings can result in severe injury or death.
• Read the instructions to ensure safe operation.
• Initial calibration should be performed in an area known to
be free of carbon monoxide. Calibration in an area containing
carbon monoxide will result in incorrect calibration and lower
than actual readings.
• Meter is NOT intended as personal protection equipment (PPE).
• Meter is NOT insulated. Avoid contact with areas where
energized conductive elements may be present. Shut off power
to the area before starting measurements.
• Always wear approved eye protection.
2

GENERAL SPECIFICATIONS

WARNINGS

Publicité

Table des Matières
loading

Table des Matières