Regulatory Information
† The ISM (industrial, scientific and medical) bands between 150 kHz and
80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957
MHz to 27. 283 MHz; and 40.660 MHz to 40.700 MHz.
‡ Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast, and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which the CPRmeter 2 is
used exceeds the applicable RF compliance level above, the CPRmeter 2
should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as re-orienting or
relocating the CPRmeter 2.
Recommended separation distances between portable
and mobile RF communications equipment and the
CPRmeter 2
The CPRmeter 2 is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the
user of the CPRmeter 2 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CPRmeter 2 as
recommended below, according to the maximum output power of the
communications equipment.
35