Maintenance and Cleaning
Cleaning after Training on Manikin
If the CPRmeter 2 has been used for training on a manikin, it can be
wiped using an alcohol wipe with minimum 70% ethanol.
Cleaning and Disinfection after Clinical Use
If the device has been used in a clinical situation, the device must be
cleaned and disinfected.
Cleaning
1. Prepare a cleaning solution by mixing 5 ml of a mild dishwashing
liquid in 4 l of warm tap water (40 °C – 50 °C).
2. Dip a small brush (e.g. toothbrush) in the cleaning solution and
scrub the device for a minimum of 2 minutes.
3. If the membrane port is clogged, use the wet brush to remove
any obstruction.
4. Wipe the exterior with a soft cloth dampened in lukewarm
water (22 °C - 40 °C).
Disinfection
1. The exterior can be disinfected in a 0.55% solution of ortho-
phthalaldehyde, or a 70% solution of isopropyl alcohol
(isopropanol). Spray the exterior surfaces with the solution and
ensure it remains wet for a minimum of 12 minutes. Repeat
spraying as necessary to account for evaporation.
2. Wipe the exterior at least 3 times with a clean damp cloth to
remove all traces of the disinfectant.
3. Allow to dry completely.