RT-3, RT-5x, RT-7x and RT-9x Thermal Imagers
bient conditions does not change the thermal im-
ager output.
1. Go to Menu, Configuration, Ambient condi-
tions.
2. Move the joystick as necessary to set the val-
ues.
3. Press OK.

Information

The following instrument information is displayed:
• Device data (e.g. serial number, device name,
firmware version)
• Options
• Measurement functions
• Wi-Fi
• Radio certifications
• Legal information

Reset Options

Reset Counters
After a reset, the naming (consecutive
NOTI E
numbering) of the images starts from the begin-
ning. The already saved images with the same
name will be overwritten. Back up all saved im-
ages before resetting the counter to prevent pos-
sible overwriting.
1. Open Menu and press ok.
2. Move the joystick up/down until configuration
is selected.
3. Move the joystick to the right, select Reset
counter.
4. Reset image counter? is displayed on the
screen.
5. Press OK to apply factory settings.
6. Press Esc to cancel the process.
Factory Settings
The instrument settings can be reset to the fac-
tory settings. Time/date, country settings and
counter are not reset.
1. Open Menu and press ok.
2. Move the joystick up/down until configuration
is selected.
3. Move the joystick to the right, select Reset op-
tions.
4. Move the joystick up/down to select the Fac-
tory settings function.
16
5. Apply factory settings? is displayed on the
screen.
6. Press OK to apply factory settings.
7 . Press Esc to cancel the process.
Formatting
The image memory can be formatted to remove
all images from the imager. When formatting, all
data saved in the memory is lost. Back up all
saved images before formatting to prevent loss of
data. Formatting does not reset the counter.
1. Open Menu and press ok.
2. Move the joystick up/down until configuration
is selected.
3. Move the joystick to the right, select Reset op-
tions.
4. Move the joystick up/down to select the For-
mat function.
5. Format memory? is displayed on the screen.
6. Press OK to format the memory.
7 . Press Esc to cancel the process.

Storage

Remove battery from tool. Store the Thermal Im-
agers in case. Avoid storing in extreme heat or
cold.
Store tool in a dry, secured area
WARNING
that is out of reach of children and people un-
familiar with the RIDGID Thermal Imagers.
The thermal imagers are dangerous in the
hands of untrained users.
Maintenance
WARNING
Remove battery from tool before clean-
ing.

Cleaning

Do not immerse the thermal imagers in water.
Close the battery cover and USB port cover.
Wipe off dirt with a damp soft cloth. Avoid rub-
bing too hard. Do not use aggressive cleaning
agents or solutions.
Gently clean the display screen with a clean dry
cloth. Avoid rubbing too hard.
If the infrared camera lens is dirty, clean it with
an alcohol lens wipe. Treat the instrument as you
would a telescope or camera.
999-995-119.10_REV. A