Dc And Ac Voltages (V) - Chauvin Arnoux C.A 5230 G Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
3
FUNCTION SELECTOR SWITCH
OFF
Off position
~
V
AC voltages up to 600 V in rms value
V ... —
DC voltages up to 600 V
Hz
Frequencies of voltages up to 400 kHz
Continuity sound test for a resistance R - 40 Ω, and
measurements of resistances up to 40 MΩ
NB: The continuity test
instrument is switched on
Diode test. Measurement of the voltage of the junction in
forward direction
Capacitances up to 40 µF
mA
Current ranges up to 400 mA DC and AC (via the + terminal)
A
10 A DC and AC current ranges (via the 10 A terminal)
4
TERMINALS
Ø 4 mm safety terminals
COM : common, terminal receiving the black lead
+ : terminal receiving the red lead
10A : terminal receiving the red lead for the 10 A DC and AC ranges
3 - DC AND AC VOLTAGES
Connect the leads to the multimeter and connect it in parallel to
the circuit to be tested.
Place the selector switch on the voltmeter function V AC or V DC
Automatic selection of the range: read the value measured.
Memorise the value, if necessary, by pressing the HOLD button.
If necessary, press the following buttons:
RANGE to select the ranges manually
MIN-MAX to read the min or max values
to light the display
Frequencies : 40 Hz – 500 Hz, except 400 mV device (40 Hz – 100 Hz)
V DC and AC
400 mV
(1)
Digital
0.1 mV
1 mV
resolution
Bargraph
10 mV
100 mV
resolution
Impedance
100 MΩ
Accuracy
on DC
Accuracy
± 1.5% rdg
on AC
± 5 counts
Permitted
overload
(1) Accessible by the RANGE button. Min measurement: 20 mV
(2) Crest factor FC - 3
is selected automatically when the
4V
40 V
400 V
10 mV
100 mV
1 V
10 V
10 MΩ
9.1 MΩ
± 0.25% rdg ± 1 count
± 1% rdg ± 3 counts
600 V rms and 1000 V peak
600 V
(2)
1 V
100 V
15

Publicité

Table des Matières
loading

Table des Matières