SECTION 4—COMPRESSOR OPERATION
SECTION 4—COMPRESSOR
OPERATION
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
NEVER use tools of any kind to connect/disconnect the
cylinder and the compressor. Otherwise, severe injury
and/or damage may occur.
DO NOT drop oxygen cylinders. Use both hands when
handling/transporting oxygen cylinders. Otherwise, injury
or damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE: For this procedure, refer to FIGURE 4.1 on page 23 and
FIGURE 4.2 on page 24.
1. Examine the cylinder gauge. If the cylinder pressure is less
than 1500 psig (10342 kPa), proceed to STEP 2 (FIGURE 4.2).
NOTE: If the cylinder pressure is greater than 1500 psig (10342
kPa), DO NOT attempt to top off this cylinder. It may not fill.
2. Set the regulator flow dial on the cylinder to Off (FIGURE
4.2).
3. Remove the cylinder and connector fillport covers.
DO NOT connect the outlet (cannula) connection to the
HomeFill II unit as this will result in the malfunction of the
conserving cylinder.
HomeFill
II Compressor
®
WARNING
CAUTION
22
Part No 1145804