5. Press the Sample key. [------] will be displayed in both
the piece window and the count window.
6. The established APW and count will be displayed.
ENTERING A KNOWN APW
1.
Press the APW key. [------] will be displayed in the
piece window.
2. Input the APW value.
3. Press the APW key. The new APW value will be
displayed in piece weight window.
SIMPLE COUNTING
1. After a valid APW is established, place a container
on the pan and press Tare if required.
2. Place the sample on the pan.
3. The quantity will be displayed in the count window.

ACCUMULATION AND STATISTICS

The Accumulation feature enables manual or
automatic totalizing of displayed values. Statistical
data (accumulated weight, count and number of
weighments) is stored in memory for review and printing.
Accumulation works together with Check mode.
ACCUMULATING DISPLAYED VALUES
With ACCUMULATE set to MANUAL, place the item on
the scale and press the M+ key to add the weight to
accumulation data. The
the weight is removed.
With ACCUMULATE set to AUTO, place the item on
the scale. The displayed value is accumulated
automatically. The
pointer will keep flashing until the
weight is removed.
VIEWING AND CLEARING STATISTICAL DATA
When the pan is cleared, press the M+ key to display
the statistical information. To clear the accumulation
data, press the Tare key while the statistical information
is displayed. The display shows [Clr.aCC]. Press the Yes key
to clear the stored data and return to current mode.
NOTE: The item must be removed from the pan
before the next item can be accumulated.
NOTE: Only stable weights are stored to
accumulation data.
NOTE: Changing the mode will clear the
accumulation data.
PAGE 6 OF 69
OPERATION CONTINUED
pointer will keep flashing until
NOTE: When Legal for Trade is on, the display
must return to 0 gross. Otherwise, the
will continue flashing. Gross loads and net
loads cannot be added to the same total. If
the first load is a gross weight, future loads
must also be gross weights. If the first load is
a net weight, future loads must also be net
weights.

CHECK-WEIGHING

Press and hold the Target key. [ChEck] is displayed in the
weight window. Release the Target key when [WEIGH] is
displayed in the piece weight window. Use this mode to
compare the weight of items to a target weight range.
1.
Press the Target key. [UndEr] is displayed in the
weight screen. The previous under limit is displayed
in the piece weight screen.
2. Input the target weight with the numeric keypad. The
value will be displayed in the count window.
3. Press the Yes key to accept the under limit value.
[OVEr] is displayed in the weight screen.
4. Repeat steps 2 and 3 above to set the over limit.
5. If required, place an empty container on the pan
and press Tare.
6. Place sample material on the pan or in the
container. If the sample weight is under the target
weight range, the yellow LED will light. If the sample
is within the target weight range, the green LED will
light. If the sample is over the target weight range,
the red LED will light.

CHECK-COUNTING

Press and hold the Target key. [ChEck] is displayed in the
weight window. Release the Target key when [Count] is
displayed in the piece weight window.
1.
Press the Target key. [UndEr] is displayed in the
weight screen. The previous under limit is displayed
in the piece weight screen.
2. Input the sample size with the numeric keypad. The
value will be displayed in the count window.
3. Press the Yes key to accept the under limit value.
[OVEr] is displayed in the weight screen.
4. Repeat steps 2 and 3 above to set the over limit.
5. If required, place an empty container on the pan
and press Tare.
pointer