CLEANING
The housing may be cleaned with a cloth dampened
with a mild detergent if necessary. Do not use solvents,
chemicals, alcohol, ammonia or abrasives to clean the
housing or control panel.
SYMPTOM
Cannot turn on
Poor accuracy
Cannot calibrate
Cannot access mode
Cannot access unit
Battery icon flashing
Err 8.1
Err 8.2
Err 8.3
Err 8.4
Err 8.5
Err 8.6
Err 9.5
-------
--NO--
CAL E
Lo.AEF
rEF.Err
NO.ACC
Battery fails to charge fully

SERVICE INFORMATION

If the troubleshooting section does not resolve or
describe problem, contact Uline Customer Service at
1-800-295-5510.
PAGE 18 OF 69
MAINTENANCE
POSSIBLE CAUSE
No power to scale
Improper calibration
Unstable environment
Unstable environment
Incorrect calibration weight
Mode not enabled
Unit not enabled
Battery is empty
Power on error
Power on error
Over range error
Under range error
Tare out of range
Display overflow
Calibration data error
Busy
Action not allowed
Calibration error
Unstable environment
Incorrect calibration weight
Low reference weight warning
Unacceptable reference weight
Gross and net weights cannot be accumulated
together
Battery is defective
TROUBLESHOOTING
Table below lists common problems and possible
causes and remedies.
If the problem persists, contact contact Uline Customer
Service at 1-800-295-5510.
Verify connections and voltage.
Perform calibration.
Move scale to suitable location
Move the scale to suitable location.
Use correct calibration weight.
Enter menu and enable mode.
Enter menu and enable unit.
Connect the scale to AC power and charge the battery.
Weight reading exceeds Power On Zero limit.
Weight reading below Power On Zero limit.
Weight reading exceeds Overload limit.
Weight reading below Underload limit.
Adjust tare value to be within range.
Weight exceeds six digits.
Calibration data not present.
Displayed during tare setting, zero setting and printing.
Function not executed.
Calibration value outside allowed limits.
Move the scale to suitable location.
Use correct calibration weight.
Increase reference weight.
Reference weight too small.
Weight on the pan is too small to define a valid
reference weight. Increase reference weight.
Only accumulate gross or net weights
(only works in LFT).
Call Uline Customer Service at 1-800-295-5510 to
replace battery.
REMEDY
1021 IH-9554