AUTO TARE [A.TARE]
Set the automatic tare functionality.
• OFF = Automatic Tare is disabled.
• ON = The first stable gross weight is tared.
• ON-ACC = Stable gross loads within the accept
limits are tared (in Check weighing mode).
BEEPER OPT [BP.OPT]
Set whether or not the beeper sounds when APW has
been optimized.
• OFF = disabled
• ON = enabled
BEEPER SIGNAL [BP.SIG]
Set how the beeper responds in Check Weighing mode.
• OFF = The beeper is disabled.
• ACCEPT = The beeper will sound when the weight is
within the accept range.
• UNDER = The beeper will sound when the weight is
below the under setting.
• OVER = The beeper will sound when the weight is
above the over setting.
• OVER-UNDER = The beeper will sound when the
weight is below the Under setting or above the over
setting.
BEEPER KEY [BP.KEY]
Set whether the beeper sounds when a button is
pressed.
• OFF = disabled
• ON = enabled
ACCUMULATION [ACCUMM]
Set the accumulation functionality.
• OFF = accumulation disabled
• AUTO = automatic accumulation
• MAN = manual accumulation
END SETUP [END]
Advance to the next menu or return to the top of the
current menu.
PAGE 10 OF 69
MENU SETTINGS CONTINUED
READOUT MENU
Enter this menu to set user preferences. Default settings
are in bold.
Reset
Stable Range
Filter
Auto-Zero Tracking
Light
Auto off
End Readout
RESET [RESET]
Reset the Read menu to factory defaults.
• NO = not reset
• YES = reset
STABLE RANGE [STABLE]
Set the amount the reading can vary while the stability
symbol remains on.
• 0.5d = 0.5 scale division
• 1d = 1 scale division
• 2d = 2 scale division
• 5d = 5 scale division
FILTER [FILTER]
Set the amount of signal filtering.
• LOW = less stability, faster stabilization time
• MED = normal stability, stabilization time
• HI = greater stability, slower stabilization time
AZT [AZT]
Set the automatic zero tracking functionality.
• OFF = disabled
• 0.5d = The display will maintain zero until a change
of 0.5 divisions per second has been exceeded.
• 1d = The display will maintain zero until a change of
1 division per second has been exceeded.
• 3d = The display will maintain zero until a change of
three divisions per second has been exceeded.
no, yes
0.5, 1, 2, 5
low, medium, high
off, 0.5, 1, 3
off, on, auto
off, 1, 5, 10
Exit menu
1021 IH-9554