Addendum to IFU SmartLite FOCUS – Compact Pen-Style, LED Curing Light
Guidance and manufacturer´s declaration as of 2013-06-23
Guidance and manufacturer´s declaration – electromagnetic immunity
Table 202 according to EN 60601-1-2 : 2001+A1:2006
The SmartLite FOCUS is intended for use in the electromagnetic environment specified below. The customer or the
SmartLite FOCUS should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient /
bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 HZ) magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
2 / 4
IEC 60601-
test level
± 3 kV
± 3 kV
contact
contact
± 8 kV
± 8 kV
air
air
± 2 kV
Not applicable
for power supply lines
± 1 kV
for input/output lines
± 1 kV
Not applicable
line(s) to line(s)
± 2 kV
line(s) to earth
Not applicable
<5 % U
T
for 0,5 cycle
(>95% dip in U
)
T
40 % U
T
for 5 cycles
(60 % dip in U
)
T
70 % U
T
for 25 cycles
(30 % dip in U
)
T
< 5 % U
T
for 5 s
(>95 % dip in U
)
T
3 A/m
Not applicable
user of
Compliance level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electromagnetic environment –