Advanced Bionics has performed two clinical studies of
AutoSense OS on the Naída CI M90 sound processor:
a confirmatory study comparing a Naída CI M sound
•
processor to a Naída CI Q sound processor with 10
subjects using T-Mics and electrical stimulation only
•
a supplemental study comparing AutoSense OS
On with AutoSense OS Off with 10 subjects using
a Naída CI M sound processor with either the
M Acoustic Earhook or the M T-Mic
Confirmatory Study
A total of 10 subjects were enrolled in a prospective,
confirmatory study using a within-subjects repeated-
measures design where each subject served as his/her
own control. All 10 subjects used the used the M T-Mic
and electrical stimulation only. The data presented
below show the comparison between AutoSense OS
on a Naída CI M sound processor to AutoSound on
a Naída CI Q sound processor in quiet and noise as well
as a comparison of AutoSense OS On vs AutoSense OS
Off on a Naída CI M sound processor in noise.
Sentences in Quiet at 65 dBA
Similar sentence recognition scores in quiet were
observed between AutoSense OS on a Naída CI M
sound processor and AutoSound on a Naída CI Q sound
processor. See the table below.
38