Organon TOF-Watch SX Manuel D'utilisation page 17

Dispositif destiné au monitorage du bloc neuromusculaire en salle d'opérations ou de réanimation
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

The following parameters can be changed in the set-up menu: Default settings are shown bold.
Set-up parameter:
TOF s flashing (TOF s repetition time)
TOF s repetition time can be adjusted between 1 and
I
3
:
60 minutes.
Surface electrode + stimulation flashing (stimulation units)
mA:
Surface electrode stimulation strength in milli-ampères.
µC:
Surface electrode stimulation strength in micro-coulomb.
Surface electrode + stimulation flashing (pulse width)
200 µs:
Default surface stimulation pulse width can be selected
to 200 or 300µs.
Surface electrode + stimulation flashing (stimulation size)
50 mA:
Default surface electrode stimulation strength can be
adjusted between 0 and 60 mA (0 and 12/18 µC).
Needle electrode + stimulation flashing (stimulation units)
µC:
Needle stimulation strength shown in micro-coulomb.
mA: Needle stimulation strength shown in milli-ampères.
Needle electrode + stimulation flashing (stimulation size)
0.0 µC:
Default needle electrode stimulation strength can be
adjusted between 0.0 and 6.0 mA/0.24 µC.
P flashing (P button function)
- - :
no function assigned to P button
DBS 3.3:
DBS 3.3 stimulation assigned to P button
DBS 3.2:
DBS 3.2 stimulation assigned to P button
50 Hz:
50 Hz tetanic stimulation assigned to P button
100 Hz:
100 Hz tetanic stimulation assigned to P button
Lo + loudspeaker + transducer flashing (Low TOF alarm)
- - :
Low TOF alarm not active
1-4:
Low TOF alarm will sound if the number of TOF
responses are below the number of set responses.
3 - 100%:
Low TOF alarm will sound if the TOF ratio is below
set percentage.
17 | TOF-Watch SX Operator manual 33.512/A