Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 7
2. Instruct the test subject to keep the sample collector in the
cheek pouch and move it carefully from one side of the
mouth to the other (Fig. 2).
The test subject must not chew or suck on the sample
collector.
Observe the sampling process!
A sufficient sample will be collected within four minutes.
If the sample volume indicator shows blue before four
minutes have passed, the sampling can be ended (Fig. 3).
3.2

Test procedures

NOTICE
i
i
Observe the relevant Instructions for Use for the
Dräger DrugTest 5000 Analyzer.
3. Take the test cassette with the sample from the test
subject. Open the Analyzer door and insert the test
cassette into the bottom slot of the Analyzer until it audibly
clicks into position (Fig. 4).
4. Remove the cartridge from the safety cap and insert the
cartridge into the top slot of the Analyzer until it audibly
clicks into position (Fig. 5).
5. Close the door.
The Analyzer now automatically starts the analysis.
On completion of the analysis, the Analyzer displays the
results for every tested substance on the screen.
Note the display!
6. Remove the test cassette with cartridge from the Analyzer
after the analysis and dispose of it (Fig. 6).
4
Limitations of the procedure
The Dräger DrugTest 5000 TTK does not contain any
components suitable for use as drug evidence. When using
a Dräger DrugTest 5000 TTK with a red cap, the Analyzer
gives
a
positive
result;
DrugTest 5000 TTK with a yellow cap, the Analyzer gives
a negative result.
5
Storage
Store the Dräger DrugTest 5000 TTK between +4 °C and
+30 °C
(39 °F
to
86 °F).
DrugTest 5000 TTK if the foil pouch is damaged (e.g. due to a
hole or rip). Use the Dräger DrugTest 5000 TTK immediately
after opening the foil pouch.
Do not use the Dräger DrugTest 5000 TTK if the embossed
expiration date is exceeded.
6
Disposal
Dräger DrugTest 5000 TTKs can be disposed with normal
domestic waste.
6
when
using
a
Dräger
Do
not
use
the
Dräger
en
Dräger DrugTest 5000 TTK

Publicité

Table des Matières
loading

Table des Matières