Télécharger Imprimer la page

Troubleshooting; Printer Settings - Mettler Toledo P25 Mode D'emploi

Masquer les pouces Voir aussi pour P25:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

7

Troubleshooting

Problem / Message
Printout unreadable
Pilot lamp does not light
P
ilot lamp flashes
---Bauderate not found---
*Please connect device and
switch the printer off/on.
*Alternatively you can change
the settings of the printer
or the connected device.

Printer settings:

Baudrate
Bit/Parity
8 Bit None
Handshake
--------Timeout ---------
------- Over Load -------
------ Under Load -------
-- Transmission Error --
---- Not Executable ----
--- Date/Time Not Set --
----- Memory Error -----
----- Memory full ------
-- Load/Unload Weight --
16
Printer P25, P26, P28
Possible cause
• Ribbon worn or entangled.
• Service life of printing unit
at an end.
• No power.
• Power supply fault.
• No connection printer
device.
• Printer did not find the
settings of the connected
instrument.
9600
Off
• Data transfer has not
completed in the last 45 s.
• Balance in overload range.
• Zeroing out of the zero
range.
• Balance in underload
range. Weighing pan is not
in position.
• Instrument does not under-
stand printer instruction.
• Instrument can not execute
printer instruction.
• Printer error messages.
• Statistics or Totaling
Memory is full.
• No sample or the same
sample is on the weighing
pan.
Correction
• Change ribbon cartridge or
tense the ribbon properly.
• Please contact MEETLER
TOLEDO service.
• Switch printer on.
• Please contact METTLER
TOLEDO service.
• Make sure that the printer and
the instrument is connected.
• Switch the printer off/on.
• Alternatively change the
settings of the printer or the
connected instrument.
(see chapter 8)
• Move balance to more stable
location.
• Check the balance.
• Check the weight of sample.
• Change the zero range of the
balance.
• Change the setting of the
printer from zero to tare.
• Check the balance.
• Check interface parameter
settings on printer and instru-
ment.
• Check the balance.
• Check the instrument.
• Reset the printer.
• Set date/time.
• Back-up battery depleted
Please contact METTLER
TOLEDO service.
• Start a new Statistics or Total-
ing application.
• Put a new sample or remove
it and put it again on the pan.

Hide quick links:

Publicité

loading

Ce manuel est également adapté pour:

P26P28