Specifi cations
Input Data for Thermocouples (TC)
Sensor type (Standard)
J
(DIN 60584-1)
K
(DIN 60584-1)
Input resistance
Max. line resistance
Line monitoring
Accuracy
Temperature coeffi cient at the
input
Cold junction compensation
Error of external
cold junction compensation
Error of internal
cold junction compensation
44
Range [°C]
- 210 ... 1200
- 200 ... 1372
> 10 MΩ
1 kΩ
Open circuits
± (10 μV + 0.05 % meas.val.)
for spans > 2 mV
50 ppm/K of adjusted end value
(average TC in permitted operating
temp range, reference temp 23 °C)
Internal or external
Via Pt100 for T comp = 0 ... 80 °C:
± (80 mΩ + 0.1% meas.val.)
± 1.5 °C