Diagnostics - Troubleshooting - Breakdowns - Tecno-gaz PURA plus 249S2 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ENGLISH
DIAGNOSTICS – TROUBLESHOOTING - BREAKDOWNS
TROUBLESHOOTING TABLE
ATTENTION: WITH AUTOCLAVE SWITCH SWITCHED ON
Anomaly
Cause
st
1
Autoclave is
not requesting
water
nd
2
Supply tap
PURA
closed
PLUS™
produces
no water
rd
3
No power
st
1
Low
waterworks
PURA
pressure
PLUS™
nd
2
Defective or
produces
spent osmotic
very little
membrane
water
The examples mentioned above are typical anomalies that may occur. Everything
outside the third column pertains to Tecno-Gaz S.p.a.
Normal solution
(possibly adopted by the user)
Check that the max level led of the
autoclave is off (see PAR 8).
Check that the tap supplying PURA
PLUS™ is not closed by mistake.
Check that you have inserted the
connector correctly into the
autoclave
Check that the waterworks
pressure is not below 2 bar.
NA
Contact the Service centre.
16
Solution by TECNO-
GAZ authorized
technicians
-
-
Check that the
24VAC on the
contact 1 e 2, in the
autoclave, is
present.
Find the cause of
the low pressure
and eliminate it.
Replace the osmotic
membrane.

Publicité

Table des Matières
loading

Table des Matières