ENGLISH
OPERATORS ISTRUCTIONS
CONNECT PURA PLUS™ ONLY TO AUTOCLAVES COMPATIBLE WITH THAT
UNIT.
ALWAYS INSTALL A TAP UPSTREAM OF THE DEMINERALIZATION CIRCUIT
IN ORDER TO AVOID FLOODS;
AT THE END OF THE DAY OR WHEN THE AUTOCLAVE IS NOT GOING TO
BE USED FOR SOME DAYS, CLOSE THE TAP INSTALLED UPSTREAM OF
PURA PLUS™
CONNECTING PURA PLUS™ TO THE HOT WATER LINE IS STRICTLY
FORBIDDEN!
PURA PLUS™ CAN BE MOUNTED TO THE WALL OR CAN BE SUPPORTED
ON A SURFACE.
THERE IS NO MINIMUM CAPACITY. THE MAXIMUM CAPACITY IS 12 l/h
MAXIMUM AND MINIMUM PRESSURE FOR WATER CLEAN OPERATION:
from 2 to 7 bar
USE OF PURA PLUS™ IS GRENTED WITH MAX HARDNESS OF THE
INCOMING WATER OF 200 °F.
OUTPUT CONDUCTIVITY VALUE: < 15 µS (Micro Siemens); THE INPUT
HARDNESS IS IRRELEVANT.
THE MEMBRANE AND RESINS NEUTRALIZE ALL KINDS OF ORGANIC AND
INORGANIC SALT.
THE CARTRIDGES MUST BE HANDLED WITH NORMAL PRECAUTIONS. THE
USE OF PROTECTIVE GLOVES IS MANDATORY
THE FILTERING CARTRIDGES MUST BE DISPOSED OF AS SPECIAL WASTE
(SEE SEC.7)
INSTALL PURA PLUS™ WITHIN 2 METRE OF THE AUTOCLAVE
DO NOT EXPOSE PURA PLUS™ THE WATER CLEAN TO FROST
DO NOT INSTALL NEAR SOURCES OF HEAT
DO NOT INSTALL UNDER OTHER TAPS
14