bit 3 =
Detection
bit 2 =
bit 1 =
bit 0 =
Example: To return the RQS bit (bit 6 of the Status Byte Register) as a 1 when true
and a 0 when false in bit position 1 in response to a parallel poll operation
send the following commands
*PRE 64 <pmt>, then PPC followed by 69H (PPE)
The parallel poll response from the generator will then be 00H if RQS is 0
and 01H if RQS is 1.
During parallel poll response the DIO interface lines are resistively terminated
(passive termination). This allows multiple devices to share the same response bit
position in either wired-AND or wired-OR configuration, see IEEE 488.1 for more
information.
5.5 Status Reporting
This section describes the complete status model of the instrument. Note that
some registers are specific to the GPIB section of the instrument and are of limited use
in an RS232 environment.
5.5.1 Standard Event Status and Standard Event Status Enable Registers
These two registers are implemented as required by the IEEE std. 488.2.
Any bits set in the Standard Event Status Register which correspond to bits set
in the Standard Event Status Enable Register will cause the ESB bit to be set in the
Status Byte Register.
The Standard Event Status Register is read and cleared by the *ESR?
command. The Standard Event Status Enable register is set by the *ESE <nrf>
command and read by the *ESE? command.
Bit 7 -
Power On. Set when power is first applied to the instrument.
Bit 6 -
Not used.
Bit 5 -
Command Error. Set when a syntax type error is detected in a
command from the bus. The parser is reset and parsing continues at
the next byte in the input stream.
Bit 4 -
Execution Error. Set when an error is encountered while attempting to
execute a completely parsed command. The appropriate error
number will be reported in the Execution Error Register.
Bit 3 -
Not used.
Page 24
Sense of the response bit; 0 = low, 1 = high
?
?
Bit position of the response
?
USER'S MANUAL. GR-205