Transcend 3 minicpap Guide Rapide page 15

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17
Transcend 3 miniCPAP Quick Guide
Cleaning of Accessories
The following accessory should be cleaned with a 5% solution of mild liquid detergent in
distilled water (1.6 fl oz liquid cleaning detergent per quart of distilled water). Mild
detergent should contain biodegradable anionic surfactants and no phosphate. Follow
these steps to clean the accessories.
Accessory
Air Supply Tube
1.
Fully immerse the air supply tube in the cleaning solution.
2.
While immersed, thoroughly wipe the surface with a lint-free cotton cloth. Apply
firm pressure and ensure contact with all accessible contact surfaces to adequately
remove soil buildup.
3.
Clean the inside of the air supply tube by lifting, then lowering the ends of the tube
while the tube is filled with cleaning solution.
4.
Rinse air supply tube by immersing in distilled water. Move the air supply tube in a
back and forth motion for approximately10 seconds to remove cleaning agent
residue.
5.
Rinse the air supply tube in distilled water by fully immersing. Lift, then lower, the
ends of the tube while the tube is filled with water. Repeat this motion for
approximately 10 seconds to remove cleaning agent residue.
6.
Dry the outside of the air supply tube with a dry, lint-free cotton cloth. Allow the
tube to air dry until the inside of the tube is dry. Length of drying time will depend
on ambient conditions.
Questions? Contact Somnetics customer service 877-621-9626 or
Periodic Cleaning Cycle
Daily
info@somnetics.com.
Product Service Life
3-Year
Page 15

Publicité

Table des Matières
loading

Table des Matières