7. CARE AND MAINTENANCE
Routine inspection listed below shall be performed every day for the sake of safety.
1. Check if there is any damage or wear by visual inspection.
2. Make sure screws and nuts on lifting arm are secured.
3. Check if hook on lifting arm works properly.
4. Controller, actuator and base width adjustment verification.
5. Emergency descend function verification (both manually and electrically).
If needed, to clean the patient lift with warm water or alcohol. Make sure there is no dust and hair on the wheels.
Atenttion: to clean the patient lift with chemicals with Phenol and chlorine is prohibited, which may cause damage on plastic powder
or plastic material.
Transportation and Storage:
Press the emergency stop button for transportation or not using the patient life for a certain of time. Temperature for storage of patient
life shall be higher than 0ºC and relative moisture ≤60%.
Inspection:
Make a full inspection at least once a year. Pay close attention on those worn part such as connection and screw bolt.
8. TROUBLE SHOOTING
DESCRIPTION
Patient lift cannot work
(ascend and descend)
Charger cannot charge the battery
Patient lift stops when ascending
16
Electric Patient Lift FORTUNA
EN
ES
DESCRIPTION
1. Check if the emergency stop button is pressed
down.
2. Verify the controller and cables from actuator is
Patient lift cannot work
properly connected
(ascend and descend)
3. Check the battery level from display screen on
the control box.
4. Verify if there is any problem or damage on
battery contact panel
.
1. Check if the charger fully plugs into socket
2. Verify if there is any problem or damage on
Charger cannot charge the battery
battery contact panel.
1. Check if the emergency stop button is pressed
Patient lift stops when ascending
down.
2. Employ the emergency descend device (see
user guide)
TROUBLE SHOOTING