Chemically-resistant Laboratory pump N820.18 EX / N840.18 EX
Fig.4: Operating
button for gas bal-
last
Translation of Original Operating Instruction, KNF 317428-317425 05/21
7.1 Gas ballast
DANGER
If vaporous media are transferred, the formation of con-
densate in the pump heads can be minimized by open-
ing the gas ballast valve.
The final vacuum that can be achieved is worse when
the gas ballast valve is open (see the chapter entitled 6
Technical data [} 20]).
Personal injury through poisoning or
explosion and damage to the pump
à
When the gas ballast valve is open,
make sure that no reactive or ex-
plosive mixtures can form.
à
Close the gas ballast valve if nec-
essary.
à
If inert gas is necessary, contact
KNF Service.
Design and function
25