Pump Shutdown; Power Failure; Remote Control Mode Operation; Rs 232 Control Mode Operation - Varian Turbo-V 150 PCB Mode D'emploi

Table des Matières

Publicité

With normal LOW SPEED operation, the pump will run
at about 2/3 of its nominal speed and achieves a base
pressure
somewhat
specifications. If the gas load becomes higher, the
controller output frequency and voltage start to
decrease automatically, and the Turbo-V pump is
slowed down in proportion to the gas load until it
reaches about 3 KRPM.
If the LOW SPEED mode is selected after normal
operating condition is reached, the display shows:
A P P R O A C H
X X
while approaching the low speed value.
When the low speed mode is deselected, the pump
starts to accelerate to its rotational speed. When this is
reached, the display shows:
N O R M A L
X X

Pump Shutdown

Press the front panel STOP pushbutton or remove the
remote signal; the power from the turbopump will be
removed and the pump will begin to slow down.

Power Failure

In the event of a power failure (momentary or long
term), the Turbo-V controller will stop the turbopump
and all the interconnected pumps/devices. The Turbo-V
vent valve device, if used, will vent the turbopump only
if the power failure is longer than the preset delay time.
When power is restored, the Turbo-V controller
automatically restarts the interconnected devices and
the turbopump in the proper sequence.
The display shows:
P U M P
I
S
1 2
X X
until normal operation achieved.
higher
than
the
standard
I
N G
L S
K R P M
O P E R A T
I O N
K R P M
S T A R T
I
K R P M

Remote Control Mode Operation

If remote signals are used to operate the controller, it
must be programmed for remote operation (see
paragraph
when ready to start, the display shows:
R E A D Y
S O F T
If the Soft Start has been deselected the display shows:
L S
P U M P
R E M O T E
With or
START/STOP
pushbuttons
NUMBER and PUMP CURRENT pushbuttons are
always active.

RS 232 Control Mode Operation

If the RS 232 option is installed and the controller has
been programmed for RS 232 operation, the controller
may be driven by a computer and when ready to
operate, the display shows:
R E A D Y
S O F T
If the Soft Start has been deselected the display shows:
P U M P
R S 2 3 2
With or
START/STOP, LOW SPEED functions are under
N G
computer control, while the CYCLE NUMBER and
PUMP CURRENT front panel pushbuttons are always
active.
68
TECHNICAL INFORMATION
"FRONT/REMOTE/232
F O R
R E M O T E
S T A R T
R E A D Y
S T A R T
wihout
Soft
Start
mode
and
LOW
SPEED
are
inoperative,
while
F O R
R S 2 3 2
S T A R T
R E A D Y
L
I
N E
wihout
Soft
Start
mode
87-900-908-01 (C)
Selection")
and
:
U S E
selected
the
front
panel
the
CYCLE
:
U S E
selected
the