AbraPlan-10
Instruction Manual
AbraPlan-10
Safety Precaution Sheet
To be read carefully before use
1.
The operator should be fully aware of the use of the machine according
to the Instruction Manual.
2.
The machine must be installed in compliance with local safety
regulations.
3.
Be aware that the machine's centre of gravity is located in the upper
half of the machine.
4.
Before lifting the machine, ensure that the supplied lifting boom is
securely bolted to the machine.
5.
When lifting the machine using a forklift, lift from front or rear - never lift
the machine from the side.
6.
When lifting the machine using lifting straps, ensure that the straps are
crossed and do not press on the sides of the machine.
7.
The machine must be placed on a safe and stable support, which is
capable of bearing the weight of this machine. Before using the
machine, it must be levelled using the adjustable legs.
8.
Be sure that the actual voltage corresponds to the voltage stated on the
side of the machine and that the connections comply with local
regulations. The machine must be earthed.
9.
Be aware that when the machine is connected to a compressed air
supply the specimen holder arm moves upwards.
10. Make sure that the specimens in the specimen holder are securely
fixed, and ensure that the securing screws are not sticking out.
11. Keep clear of the grinding stone when operating.